- With solid membrane Interchangeable flat surface
- Ion to be detected passes through the membrane generating a difference of potential, proportional to ionic concentration
- Analysis time is generally less than one minute
- Sample preparation steps such as filtering or Distillation rarely necessary
- Calibration should be carried out as often as possible (once a day) at the same temperature as the sample
- Measurement temperature 0 to 50°C
- Delivered with membrane, electrolytic solution, ISAB solution and ISE standard and gasket
ISAB solution:
To adjust the ionic strength. add directly to the solution to be measured.
ISE standard:
For the calibration of ionic electrodes. Concentration 1000 ppm.
Especificaciones
Especificaciones
| Applications | Determination of NO3- ion |
| Filling Solution Type | ISAB/ISE |
| pH Range | 2.5 to 11 |
| Interferences | CIO4-, I-, CN-, BF4-, Cl-, NO2- |
| Electrode | Nitrate Combined ISE |
| Parameter | Nitrate Ion |
Título del producto
Al hacer clic en Enviar, acepta que Fisher Scientific se ponga en contacto con usted en relación con los comentarios que ha proporcionado en este formulario. No compartiremos su información para ningún otro fin. Toda la información de contacto proporcionada se mantendrá de acuerdo con nuestra Política de Privacidad. Política de privacidad.